- Other Measuring & Analysing Instruments[10]
- Counters[1]
- General Assay & Diagnostic Apparatus[2]
- Other Flow Measuring Instruments[1]
- Other Electrical Equipment[1]
- Temperature Instruments[3]
- Sterilization Equipments[5]
- Drying Equipment[10]
- Mixing Equipment[1]
- Cleaning Equipment[7]
- Laboratory Furniture[7]
- Other Chemical Equipment[7]
- Microscopes[7]
- Weighing Scales[2]
- Other Analysis Instruments[8]
- Contact Person : Ms. sun Lily
- Company Name : Shanghai Jingke Scientific Instrument Co., Ltd.
- Tel : 86-21-34637408
- Fax : 86-21-34637418
- Address : Shanghai,Shanghai,Room A304, No.189, Xinjunhuan Road, Minhang District
- Country/Region : China
- Zip : 201114
JK-BSC-1000IIB2 Biological safety cabinet
Biological safety cabinet
Characteristics
The vertical lamina airflow has the patterns of 100% air exhaust to the outdoor or to the air exhaust system.
HEPA and ULPA with NO LEAKING (Class 100) or ISO Class 4(Class 10)
The viewscreen of gradient of 10°make it comfortable to operate.
Non-obstructive air retum mode in the front suction channel
V notch and humanized design of the shelf can make sure the protection of negative pressure and alleviate working tiredness.
The air supply and eshaust pipelines adopt PDl Control Valve with BELIMO Actuator, which can effectively reduce air disturb fuom the outside.Interlocked light and sterilizing
Anti-splash safety spare socket and junction valve
ntelligent auto-supply system ensures the change of air volume less than 10% when the resistance of filter in creases 50% and enhance safety.Leak detection in cabinet, no leakage at 500Pa.
eak detection in HEPA/ULPA, leaking ratio is ≤0.01%(scanned),≤0.005%(not scanned)
Technical parameters:
Model | JK-BSC-1000IIB2 | JK-BSC-1300IIB2 | JK-BSC-1600IIB2 |
Cleanliness | HEPA:iso5(Class100)ULPA:ISO 4(Class 10) | ||
Fitration Efficiency | HEPA:≥99.995%,@0.3μm/ULPA: ≥99.999%,@0.12µm | ||
Downflow Velocity | 0.35 | ||
Inflow Velocity | 0.55 | ||
Noise | ≤56 | ≤58 | ≤60 |
Supply Power | AC220V,1mm,50Hz | ||
Operator protection | Total colony in impactionsampter≤10CFU/time Total colony in impactionsampter≤5CFU/time | ||
Product protection | Total colony in culture dish≤5CFU/time | ||
Cross Contamination | Total colony in culture dish≤2CFU/time | ||
Power(with spare socke) | 2.0 | 2.1 | 2.1 |
Mass | 260 | 300 | 360 |
Work dimension | 1000×600×600 | 1300×600×620 | 1600×600×620 |
Overall dimension | 1200×790×2050 | 1500×790×2050 | 1800×790×2050 |
Size and Qty of light | 36×(2) | 36×(2) | 36×(3) |
Size and Qty of UVLinght | 20×(1) | 20×(1) | 20×(1) |
Lumin | ≤800 | ≤800 | ≤800 |
Air CIRECTION | Top out |
JK-BSC-1000IIB2 Biological safety cabinet